Shasta - How to Replace a Clogged Line or Valve on a Shasta Synthesizer
This article walks through how to troubleshoot and replace any liquid valve on a Shasta system. In many cases, a valve issue can be resolved without replacement, so we recommend following the steps in order.
When should I use this guide?
Use this procedure if you see:
-
No flow during priming
-
Suspected clogging, crystallization, or valve failure
Tools & Parts You’ll Need
-
Valve fitting tool
-
Replacement valve (from Shasta spares kit)
-
Replacement tubing (if needed)
-
Empty reagent bottle
-
Gloves + eye protection
⚠️ Safety note: Always depressurize reagents before opening fittings or removing valves.
Step-by-Step Troubleshooting & Replacement
1. Check the fitting first (important)
Before assuming the valve is bad:
-
Tighten the fitting at the valve to ensure it is fully seated.
-
A loose fitting can dry out internally and restrict flow later.
-
Use the fitting tool included in your spare parts kit.
2. Try priming the valve
-
Open the Priming Screen in the Shasta software.
-
Select the affected valve.
-
Press and hold the “Hold to Prime” button.
-
Watch for flow:
-
If flow resumes, the valve was likely partially clogged and has cleared.
-
If there is still no flow, continue to the next step.
-
3. Remove liquid from the line
Before replacing any hardware:
-
Place an empty bottle in the reagent position for that valve.
-
Prime the line to pull as much liquid out as possible.
-
Depressurize the reagent system.
-
Open the bottle cap to confirm pressure is fully released.
4. Determine: tubing or valve?
At this point, the issue is usually one of the following:
A. Tubing issue
-
Look for crystallization, discoloration, or visible blockage
-
If present:
-
Remove the tubing
-
Replace it with tubing from the spares kit
-
B. Valve issue
-
If tubing looks clean and unobstructed, proceed with valve replacement
5. Replace the valve
-
Remove the two mounting screws holding the valve in place.
-
Unplug the electrical connector from the valve.
-
Plug the connector into the new valve.
-
Mount the new valve using the same two screws.
-
Snug is sufficient — do not overtighten.
-
💡 In some cases, both the valve and tubing may need replacement.
6. Re-pressurize and verify operation
-
Re-pressurize the reagent system.
-
Return to the Priming Screen.
-
Prime the valve and confirm:
-
Normal flow
-
No leaks at fittings
-
-
Once flow looks normal, resume operation.
If the issue persists
If flow still does not return after replacing both the tubing and valve:
-
Stop operation
-
Document which valve was replaced
-
Contact support with:
-
Valve position
-
Error symptoms
-
Photos if possible
-
Summary
Most Shasta valve issues are caused by:
-
Loose fittings drying out
-
Minor crystallization
-
Tubing blockages
-
Actual valve failures are less common, but replacement is quick when needed.