Skip to content
English
  • There are no suggestions because the search field is empty.

Shasta Software - Release Notes - V. 2.5.10

This article summarizes the software revisions, feature enhancements, and corrections implemented between versions v2.5.5 and v2.5.10

 

Software Release Notes


Revision History

Version 2.5.5

  • Corrected button order in the Sequence Editor — the Import and Export buttons had been reversed in a prior update.

  • Added a new configuration option to enable pre-priming all tips prior to each dispense pass (applicable to Standard and Meta protocols only).

  • Updated Run Synthesis to automatically prime all tips ready for dispense before the start of each dispense-pass series (Standard and Meta protocols only).


Version 2.5.6

  • Modified Run Synthesis pressure monitoring behavior:

    • When pressure goes out of range, the system now logs the event silently instead of displaying an interrupting dialog.

    • The synthesis automatically continues once pressure returns to the acceptable range.

    • Other pausable conditions (e.g., Lid Open, Leak Detected) will still trigger user dialogs requiring acknowledgment before resuming.


Version 2.5.7

  • Updated Backup/Restore functionality to no longer remember the previous archive path or filename.

  • Added a new configuration option to restrict plate movement unless chamber pressure is within acceptable limits.

  • Enhanced the Manual Control, Priming, and Calibrate Position screens to:

    • Display current lid switch and chamber pressure status.

    • Only allow plate motion when both lid closed and pressure OK conditions are met, based on configuration settings.

  • Updated Dispense Calibration to move the plate only when lid and pressure conditions are verified as good.


Version 2.5.8

  • Added advanced Run Synthesis controls, allowing users to:

    • Start synthesis from a specific base.

    • Pause before a specified base.

    • Pause automatically after a defined number of hours or minutes.


Version 2.5.9

  • Modified Run Synthesis to disable Pause Before Base X and Pause After X Hours/Minutes features when running MAP protocols.

 


Version 2.5.10Validated Build

  • Changed the Run Synthesis to move the plate to the "View" position at the end of the synthesis instead of a neutral (x=100, y=100) position.

Status:
Validated build for production and field use (v2.5.10)


Summary of Key Improvements

  • Enhanced pressure and motion safety interlocks for improved operator protection.

  • Added flexible synthesis controls (start/pause/resume logic).

  • Improved UI consistency across Manual Control, Priming, and Calibration screens.

  • Reduced potential for synthesis interruptions due to transient pressure fluctuations.

  • Improved pre-dispense priming reliability for Standard and Meta protocols.