Shasta Software - Release Notes - V. 2.5.10
This article summarizes the software revisions, feature enhancements, and corrections implemented between versions v2.5.5 and v2.5.10
Software Release Notes
Revision History
Version 2.5.5
-
Corrected button order in the Sequence Editor — the Import and Export buttons had been reversed in a prior update.
-
Added a new configuration option to enable pre-priming all tips prior to each dispense pass (applicable to Standard and Meta protocols only).
-
Updated Run Synthesis to automatically prime all tips ready for dispense before the start of each dispense-pass series (Standard and Meta protocols only).
Version 2.5.6
-
Modified Run Synthesis pressure monitoring behavior:
-
When pressure goes out of range, the system now logs the event silently instead of displaying an interrupting dialog.
-
The synthesis automatically continues once pressure returns to the acceptable range.
-
Other pausable conditions (e.g., Lid Open, Leak Detected) will still trigger user dialogs requiring acknowledgment before resuming.
-
Version 2.5.7
-
Updated Backup/Restore functionality to no longer remember the previous archive path or filename.
-
Added a new configuration option to restrict plate movement unless chamber pressure is within acceptable limits.
-
Enhanced the Manual Control, Priming, and Calibrate Position screens to:
-
Display current lid switch and chamber pressure status.
-
Only allow plate motion when both lid closed and pressure OK conditions are met, based on configuration settings.
-
-
Updated Dispense Calibration to move the plate only when lid and pressure conditions are verified as good.
Version 2.5.8
-
Added advanced Run Synthesis controls, allowing users to:
-
Start synthesis from a specific base.
-
Pause before a specified base.
-
Pause automatically after a defined number of hours or minutes.
-
Version 2.5.9
-
Modified Run Synthesis to disable Pause Before Base X and Pause After X Hours/Minutes features when running MAP protocols.
Version 2.5.10 — Validated Build
-
Changed the Run Synthesis to move the plate to the "View" position at the end of the synthesis instead of a neutral (x=100, y=100) position.
Status:
✅ Validated build for production and field use (v2.5.10)
Summary of Key Improvements
-
Enhanced pressure and motion safety interlocks for improved operator protection.
-
Added flexible synthesis controls (start/pause/resume logic).
-
Improved UI consistency across Manual Control, Priming, and Calibration screens.
-
Reduced potential for synthesis interruptions due to transient pressure fluctuations.
-
Improved pre-dispense priming reliability for Standard and Meta protocols.